As is known to all, the maintenance and repair of instruments and equipment must not be ignored, otherwise, it will cause the test test and other work can not be carried out smoothly.The same is true for fabric breathability meters.Below, UTS international Co., LTD engineer about some of the fabric permeability meter conventional fault discrimination and maintenance methods, hope to help you!

Conventional fault diagnosis and maintenance of fabric breathability meter


Fault 1: the device has no display when the power switch is turned on. The screen and the power display light are not on.

Possible cause: power problem, check the power supply provided.

Possible cause: a broken fuse.

To check or replace the fuse, unplug the instrument and unscrew the fuse knob to remove the fuse.


Fault 2: error range deviation occurs in validation and leak inspection.

Possible cause: check whether the hole plate of the sample is compressed and has air leakage.

Possible reasons: check whether the sample pressing device is flexible and whether the pressing ring is tight.

Possible reasons: suction hose and flow simplified, suction fan connection whether leakage.

Possible cause: the lock seal rubber ring on the flow cylinder door is leaking.


Fault 3: press the button and the instrument will not respond.

Possible cause: button broken.


General fabric breathability meter routine small fault our laboratory personnel can basically solve, if encounter procedures and other problems, must contact the manufacturer technician to solve, must not disassemble privately, so as not to cause greater losses.