The moisture permeability tester is a professional device used to measure the water vapor transmission rate (WVTR) of textiles, films, non-woven fabrics and other materials. The following are the standard operating procedures and technical points.
I. Equipment structure and principle
Core components
1. Test chamber: sealed chamber with controllable temperature and humidity (usually divided into dry/wet chamber)
2. Weighing system: high-precision balance (0.001g resolution)
3. Temperature control system: PID precise temperature control (range 20-50℃±0.5℃)
4. Humidity system: saturated salt solution or steam generator
5. Data acquisition: automatic recording of mass changes and temperature and humidity
Test principle
Positive cup method (ISO 2528): water vapor permeation from wet chamber to dry chamber
Inverted cup method (ASTM E96): water vapor absorption from dry chamber to wet chamber
Dynamic method (JIS L1099): determination of water vapor carried by airflow
II. Standard operating procedures
1. Sample preparation
Cut a circular sample with a diameter of ≥70mm (no creases/breaks)
If testing multiple layers of materials, stack them according to the actual use direction
Pretreatment: equilibrate at the test temperature and humidity for 24h (GB/T 12704 requirements)
2. Equipment initialization
1. Turn on the constant temperature water bath (set to 23℃ or 38℃)
2. Prepare saturated salt solution:
Dry environment: Mg(NO₃)₂ (RH≈53%)
High humidity environment: K₂SO₄ (RH≈97%)
3. Calibrate the balance (use standard weights)
3. Sample loading and sealing
1. Fix the sample at the mouth of the test cup:
- Positive cup method: add distilled water (liquid surface 3mm away from the sample)
- Inverted cup method: add desiccant (anhydrous CaCl₂)
2. Press with silicone seal to ensure no side leakage
3. Weigh the initial mass (m₀) to an accuracy of 0.001g
4. Test execution
1. Place the test cup in a constant temperature and humidity chamber
2. Set parameters:
Temperature: 23±1℃ or 38±1℃ (select according to the standard)
Air flow rate: 0.5-1.0m/s (dynamic method needs to be set)
Test interval: 1h/time (static method)
3. Start the test, the system automatically records the mass change (m₁, m₂...mₙ)
5. Data processing
Water vapor permeability meter:
WVTR = (Δm×24)/(A×t) Unit: g/(m²·d)
Δm: mass change (g)
A: effective test area (m²)
t: test time (h)
Example:
If the mass decreases by 0.12g within 24h and the test area is 0.00283m², then:
WVTR = (0.12×24)/0.00283 = 1017.7 g/(m²·d)